Product description
- Automatic internal adjustment in the case of a change in temperature > 2 °C and time-controlled every 4 hours.
- Dosage aid: High-stability mode and other filter settings can be selected.
- Simple weighing and documenting with a combined tare/print function. In addition, the components of the mixture are numbered automatically and printed out with their corresponding number and weight value.
- Identification number: 4 digits, freely programmable, printed on calibration protocol.
- Automatic data output to the PC/printer each time the balance stops.
Single-Cell advanced technology:
- Fully automatically manufactured weighing cell from one piece of material. Stable temperature behaviour.
- Short stabilisation time: Steady weight values within approx. 3 sec under laboratory conditions.
- Improved mechanical sturdiness.
- High corner load performance.
Reference | Weighing range (Max) g | Readout (d) mg | Verification value (e) mg | Minimum load (Min) g | Reproducibility mg | Linearity mg | Verification option |
627-ABJ80-4NM | 82 | 0.1 | 1 | 0.01 | 0.2 | 0.3 | yes |
627-ABJ120-4NM | 120 | 0.1 | 1 | 0.01 | 0.2 | 0.3 | yes |
Data sheet:
627-ABJ80-4NM (80 g):
Design | - Housing size (W x D x H): 210 mm x 340 mm x 325 mm
- Weighing area (W x D x H): 174 mm x 162 mm x 227 mm
- Housing material: Plastic
- Material of weighing plate: stainless steel
- Bubble level: yes
- Adjustable swiveling legs: yes
- Weighing surface (d): 91 mm
|
Service | - DAkkS certificate: 963-101
- Verification: 965-201
|
Display | - Height of digits: 1.400 cm
|
Verification data | - Calibration authorisation according to 90/384/EEC: yes
- Minimum load [Min]: 10 mg
- Verification class: I
- Verification value [e]: 1 mg
- Verification in factory possible: yes
|
Power supply | - Input voltage: 110 V – 230 V AC
|
Functions | - Percentage determination: yes
- Counting function: yes
- Formula function: yes
- Addition memory function: yes
- Total net function: yes
- Standby function: yes
- Weighing with tolerance range: yes
- GLP protocol: yes
|
Measurement system | - Weighing field [Max]: 82 g
- Stabilisation interval: 3 s
- Readout [d]: 0.100 mg
- Linearity: 0.300 mg
- Adjustment possibilities: Internal weight adjustment (manual) Internal weight adjustment (automatic)
- Reproducibility: 0.200 mg
- Resolution: 820
- Warm up time: 4 h
- Weighing units: g mg
|
Environmental conditions | - Maximum environmental humidity: 85 %
- Minimum environmental humidity: 20 %
- Minimum environmental temperature: 10 °C
- Maximum operative temperature: 30 °C
|
Packing | - Packing size (W x D x H): 390 x 560 x 530 mm
- Gross weight: 9 kg
- Net weight: 6 kg
|
Counting | - Minimum partial weight for piece counter: 1 mg
- Reference weight: yes
- Counting resolution: 82000
|
627-ABJ120-4NM (120 g):
Design | - Housing dimensions (W x D x H): 210 mm x 340 mm x 350 mm
- Weighing area (W x D x H): 174 mm x 162 mm x 227 mm
- Housing material: Plastic
- Material of weighing plate: stainless steel
- Bubble level: yes
- Adjustable swiveling legs: yes
- Weighing surface (d): 91 mm
|
Service | - DAkkS certificate: 963-101
- Verification: 965-201
|
Display | - Height of digits: 1.400 cm
|
Verification data | - Calibration authorisation according to 90/384/EEC: yes
- Minimum load [Min]: 10 mg
- Verification class: I
- Verification value [e]: 1 mg
- Verification in factory possible: yes
|
Power supply | - Input voltage: 110 V – 230 V AC
|
Functions | - Percentage determination: yes
- Counting function: yes
- Formula function: yes
- Addition memory function: yes
- Total net function: yes
- Standby function: yes
- Weighing with tolerance range: yes
- GLP protocol: yes
|
Measurement system | - Weighing field [Max]: 120 g
- Stabilisation interval: 3 s
- Readout [d]: 0.100 mg
- Linearity: 0.300 mg
- Adjustment possibilities: Adjustment with internal weight (automatic) Adjustment with internal weight (manual)
- Reproducibility: 0.200 mg
- Resolution: 1,200,000
- Warm up time: 8 h
- Weighing units: ct g mg
|
Environmental conditions | - Maximum environmental humidity: 85 %
- Minimum environmental humidity: 20 %
- Minimum environmental temperature: 10 °C
- Maximum operative temperature: 30 °C
|
Packing | - Packing dimensions (W x D x H): 390 mm x 563 mm x 515 mm
- Gross weight: 9.207 kg
- Net weight: 6.271 kg
|
Counting | - Minimum partial weight for piece counter: 1 mg
- Reference weight: yes
- Counting resolution: 120000
|