Product description
Veterinary anesthesia vaporiser.
- BE500 vaporiser with capacity 120 ml (isoflurane).
- It provides manual ventilation, and oxygen and anesthetic gases with a precise and constant concentration.
- It absorbs the carbon dioxide exhaled by the animal.
- The internal compliance of anesthesia breathing system shall not exceed 120 ml, at a pressure of 3 kPa (30 cm H2O).
- Handle included for easy transfer.
- Closed breathing system circuit (semi-enclosed on request).
- Two types of tubes:
- Included (22 cm) – for large animals (over 20 kg).
- Optional (17 cm) – for small animals (less than 20 kg). (Ref. 118-TB17CM)
- Animal weight:
- Minimum: 6 kg.
- Maximum:180 kg.
Technical specifications:
Gas supply:
Gas tube | O2, N2O |
Regulator output pressure | 350 kpa – 600 kPa |
Safety valve discharge pressure | ≤600 kPa |
Tube input pressure | 350 kpa – 600 kPa |
Flow meter:
Gas | Range |
O2 | 0.0 L/min ~ 10 L/min |
N2O | 0.0 L/min ~ 10 L/min |
- Accuracy: At 20ºC, the atmospheric pressure is 101.3kPa, the accuracy of the flowmeter is level 2.5.
- O2 supply: 25 L/min ~ 75 L/min
- Alarm O2 supply error: O2 pressure lower than 0.14MPa
- N2O disruption : O2 lower than 0.1 MPa
Respiratory system:
Volume | 50 ~ 1500 mL | |
Absorbent | 1200 ml × 2 | |
System leaks | When pressure is lower than 3kPa, ≤175mL/min | |
System compliance | ≤4mL/kPa | |
Vent system resistance | – Flow rate: 30L/min | – Flow rate: 60L/ kPa min |
≤0.16kPa | ≤ 0.52kPa | |
Opening pressure of unidirectional inhalation/exhalation valve | Dry status: 0.02 kPa Humid status: 0.03 kPa | |
APL valve | Over 0.19-6 kPa |
Environmental conditions:
Operation | Storage | |
Temperature | +15 ºC ~ +35 ºC | 5º C ~ +40 ºC |
Relative humidity | ≤90% non-condensing | ≤90% non-condensing |
Pressure | 96 kPa~104kPa | 50 kPa~106kPa |
Accessories included:
- Main unit
- Breathing system circuit
- Ball with reinforced connector
- O2 tube
- N2O tube
- Breathing tube
- Facial mask
- User manual
Certification:
- Directive 93/42/EEC on Medical Devices.