Hygoclave 30. Sterilisation made simple and reliable
Sterile instruments and expertise are a prerequisite for successful dental treatment. The Hygoclave 30 Class B Steam Sterilizer, provides the reassurance you need for a sustainably safe practice.

Features

  • Short heating phase thanks to matched choice of material
  • Fast cycle time from 20 min onwards (drying included)
  • Integrated fresh water quality tests
  • Protects the instruments by virtue of little thermal leaps
  • Low power consumption and excellent drying results due to a differentiated heating system
  • Automatic cleaning system
  • Diagnostic system for process evaluation and debugging
  • Elegant and functional design
  • Fast backup via USB interface and stick (Included) 
  • Optional printer

Programmes:

Programme Process parameters Load, max. load Test Type
Universal 134 ºC, 5 min
3 Pre-vacuum phases.
Drying 10 min (6 min vacuum, 4 min aeration)
All packaged and unpackaged sterile materials
up to 134 °C 4 kg solid or 1.5 kg
porous instruments and tissues, or mixed
Helix Test B
Soft 121 ºC, 20
3 Pre-vacuum phases.
Drying 11 min (7 min vacuum, 4 min aeration)
All packaged sterile materials up to 134 °C
4 kg solid or 1.5 kg porous instruments, or mixed
Helix Test B
Quick 134 ºC, 3 min
1 Pre-vacuum phases.
Drying 3 min (2 min vacuum, 1 min aeration)
Non packaged, solid sterile materials
4 kg solid instruments
S
Special P1 prions 134 ºC, 18 min
4 Pre-vacuum phases.
Drying 10 min (6 min vacuum, 4 min aeration)
Against Creutzfeld-Jakob and BSE pathogens Helix Test B
Special P2 134 ºC, 5 min
4 Pre-vacuum phases.
Drying 13 min (8 min vacuum, 5 min aeration)
All packaged sterile materials up to 134 °C
4 kg solid or 1.5 porous instruments, or mixed
Helix Test B
Bowie-Dick Test 134 ºC, 3,5 min
3 Pre-vacuum phases.
Drying 10 min (6 min vacuum, 4 min aeration)
For B&D Test (3M Comply TM, Code 1300)
B&D test pack or equivalent
B&D Test No sterilization
Vacuum Test Temperature below 35 °C Chamber empty Vacuum Test No sterilization

The Hygoclave 30 programmes can be modified at any moment and individually adapted, making it easy and fast to deal with potential changes in instrument preparation regulation. These changes are validated afterwards on site.

Technical data

  • Dimensions (L x H x W) in mm:  444 x 430 x 599
  • Chamber volume:  17.5 litres
  • Empty weight in kg:  55
  • Voltage in V:  230
  • Frequency in Hz:  50/60
  • Power in W:  1920
  • Average power in W:  1000
  • Load capacity (unpackaged instruments) in kg:

    • Solid, hollow instruments: 4
    • Porous materials: 1,5
  • Sound pressure (SPL) dB(A):  52